• Publication Date: 10/01/2026
  • Author(s):
    Czitrovszky, Aladár Farkas, Árpád Horváth, Alpár Nagy, Attila
  • Organization(s):
    Bremotech Kft. Envi-Tech Kft. HUN-REN Centre for Energy Research HUN-REN Wigner Research Centre for Physics
  • Article Type: Technical Articles
  • Subjects: Device Design, Devices and Components, Product Development/Formulation

The characteristics of air and particle flows emitted near the mouthpieces of dry powder inhalers (DPIs) are influenced not only by patient-specific factors but also by device-specific design features. The aim of this work was to design and evaluate dry powder inhaler add-on components capable of manipulating inhaled airflow to reduce upper-airway deposition and enhance pulmonary drug delivery. A further objective was to assess the suitability of the proposed add-ons across different inhaler types and patient groups.

Numerical simulations were conducted alongside experimental measurements of drug deposition and transmission. The performance and applicability of the add-ons were then evaluated based on the experimental results. The most encouraging outcome is that, for every patient’s breathing capability, at least one add-on exists that can increase the dose delivered to the lungs. At the same time, the choice of add-on must be tailored to the specific inhaler-drug combination and to the patient’s inhalation profile; otherwise, a reduction in lung dose may occur.

October 2026, Volume 20(5), pages 10-17